Automated Pressure Calibrator
CPC 6000

50

Mensor/WIKA Operating Instructions - CPC 6000

Always send commands in one of the following formats:

1.

[Command] [Termination];

2.

[Command] [Separator] [Data] [Termination];

3.

Queries are special instructions in the form:
[Command?] [Termination] where the question
mark, “?”, immediately precedes the terminator.

When a valid query is received, the CPC 6000 will return {data} terminated by CR and LF.

Floating point data is returned in the current engineering units in exponential format.

7.4.4 Output Formats

Pressure readings are returned in exponential notation in a format according to the OUTFORM com-
mand as follows:

1. pressure value
2. pressure, units number, mode
3. pressure, pressure rate
4. pressure, minimum peak, maximum peak
5. pressure, active sensor (P or S), active turndown (1-4)
6. pressure, control point, “stable” or “slewing”
7. pressure, “no barometer” or baro reading

The space at the beginning of the string is replaced with an “E” if the CPC 6000 has detected a
remote communication error.

For commands that take boolean data the following strings are acceptable:

0

1

False

True

No

Yes

Off

On