WIKA CPC6000 User Manual
Page 50

Automated Pressure Calibrator
CPC 6000
50
Mensor/WIKA Operating Instructions - CPC 6000
Always send commands in one of the following formats:
1.
[Command] [Termination];
2.
[Command] [Separator] [Data] [Termination];
3.
Queries are special instructions in the form:
[Command?] [Termination] where the question
mark, “?”, immediately precedes the terminator.
When a valid query is received, the CPC 6000 will return {data} terminated by CR and LF.
Floating point data is returned in the current engineering units in exponential format.
7.4.4 Output Formats
Pressure readings are returned in exponential notation in a format according to the OUTFORM com-
mand as follows:
1.
2.
3.
4.
5.
6.
7.
The space
remote communication error.
For commands that take boolean data the following strings are acceptable:
0
1
False
True
No
Yes
Off
On