WIKA CPC6000 User Manual
Page 61

Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
61
18
_pcs4 func stby
Instrument placed in standby mode in
19
_pcs4 func vent
Instrument placed in vent mode in
20
_pcs4 id?
Returns instrument ID.
21
_pcs4 opt?
Returns instrument options.
22
_pcs4 outform
Sets output format.
23
_pcs4 outform?
Returns the current output format.
24
_pcs4 rangemin?
Returns the minimum pressure of the active transducer.
25
_pcs4 span?
Returns the stored multiplication factor from the active
transducer and turndown.
26
_pcs4 stabledelay
1 to
255
Sets the number of consecutive readings that the pres-
sure must remain within the stable window for a pres-
sure stable indication.
27
_pcs4 stabledelay?
Returns the number of readings that must be within the
stable window before a stable pressure is indicated.
28
_pcs4 stablewindow
Sets the pressure window that is used to indicate pres-
sure is stable.
29
_pcs4 stablewindow?
Returns the pressure tolerance allowed for a stable pres-
sure indication as a % of span of the active transducer.
30
_pcs4 stat?
Returns mode and stable flag status “mode, stable CR
LF”.
31
_pcs4 unit
Sets the instrument to specified engineering units.
32
_pcs4 unit?
Returns the current engineering units and type of trans-
ducer (A, G, D).
33
_pcs4 xducer?
Returns the number of the currently active transducer.
34
_pcs4 zero?
Returns the stored zero offset of the active transducer
and turndown in the current pressure units.
7.4.7.2 PCS 200 Emulated Commands and Queries
Table 7.4.7.2 – PCS 200 Emulated Commands and Queries
Command
Data
Function / Response
CX
Control Pressure at last control point and units
C$nnnnnnnX
Control Pressure at n in units $
D#X
CAL POINT CONTROL MODE NOT SUPPORTED
EX
Clear Error/Clear Service Request
E?X
Return error code and clear error
F$nnnnnnn1X
Re-initialize; $,n ignored
F$nnnnnnn2X
RETURN CAL DATA NOT SUPPORTED
F$nnnnnnn3X
Return Unit ID string; $,n ignored
F$nnnnnnn5X
RETURN QPS TEMPERATURE NOT SUPPORTED