WIKA CPC6000 User Manual
Page 65

Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
65
:AUTOvent ON/OFF/1/0
Sets autovent function on or off (automatically vent when
controlling at zero).
:AUTOvent?
Returns status of Autovent function.
[SOURCE]
:PRESsure
[:LEVel]
[:IMMediate]
[:AMPLitude]
Sets the control pressure.
[:AMPLitude]?
Returns current control pressure value.
:SLEW
Sets the slew rate in %FS/s.
:SLEW? Returns the current slew value in %FS/s.
:TOLerance
Sets the tolerance for stable indication in %FS.
:TOLerance?
Returns the tolerance for stable indication in %FS.
CALCulate
:LIMit
:LOWer
Sets minimum control limit.
:LOWer?
Returns minimum control limit.
:UPPer
Sets maximum control limit.
:UPPer? Returns maximum control limit.
7.4.7.3.1 Output Formats
1.
2.
3.
4.
5.
6.
7.
7.4.7.3.2 Error Messages and Error Codes
A recognized command is displayed at the device in the headline with “Remote...”. In case of error
“Remote error...” is displayed. A maximum of 100 errors are stored and can be retrieved successfully.
Table 7.4.7.3.2 - SCPI Command Error Messages and Error Codes
Code
Error String Returned
0
No error
-101
Undefined character
-102
Syntax error
-103
Undefined separator
-104
Parameter data type error
-109
Parameter missing
-110
Undefined header
-113
Undefined command
-114
Parameter out of range