Automated Pressure Calibrator
CPC 6000
56
Mensor/WIKA Operating Instructions - CPC 6000
94
Rsetpt?
Returns the current rate setpoint of the active
channel in current units per second.
95
Sbaud
{9600, 19200, 38400, 57600}
Sets the serial baud rate.
96
Sbaud?
Returns the serial baud rate.
97
Sdata
{7 or 8}
Sets the serial data bits.
98
Sdata?
Returns the serial data bits number.
99
Sensor
C, X
Sets the active sensor where C = Primary or
Secondary and X is the turndown.
100 Sensor?
Returns the active sensor as above.
101 Sensor.acc?
Returns the accuracy of the sensor.
102 Sensorid?
,
No.},V{V.VV}
Returns the active sensor’s serial number and
firmware version.
103 Setpt
{A value inside the upper/low-
er limits, and inside the range
of the active transducer and
turndown}
Sets the control setpoint for the active channel.
104 Setpt?
Returns the control setpoint of the active chan-
nel in current units.
105 Span
{desired pressure or ?}
Sets span on active transducer (must be >50%
FS, and within 1% limit), or ? to clear previous
value. (See “Caldisable” command above ref#
19)
106 Span?
Returns span scale factor for active transducer
and turndown.
107 Sparity
{Even, ODD , NONE}
Sets the serial parity.
108 Sparity?
{
Returns the serial parity.
109 Srqmask
Stable, Error or both
Sets the CPC 6000 to issue a service request
(SRQ) over the IEEE-488 when the pressure
control is stable, or an error occurs. These are
80 hex and 40 hex respectively.
110 Srqmask?
Returns “stable”, “error” or “error, stable” de-
pending on the SRQ.
111 Sstop
{1 or 2}
Sets the serial stop bits.
112 Sstop?
Returns the serial stop bits.
113 Stable?
Returns YES if instrument is stable or NO.
114 Stabledelay
{0 to 65535}
Sets the stable time to the number of seconds
specified.
115 Stabledelay?
Returns the stable time.
116 Stabletime
{0 to 65535}
Sets the stable time to the number of seconds
specified.
117 Stabletime?
Returns the stable time.