WIKA CPC6000 User Manual
Page 63

Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000
63
7.4.7.3 SCPI Commands
STATus :
OPERation
:CONDition?
Returns an integer value representing instrument status
that can be decoded.
Bit 0: Zeroing active.
Bit 1: Control Setpoint has not been reached.
Bit 2: Reserved 0.
Bit 3: Reserved 0.
Bit 4: Measuring. The instrument is actively measuring.
MEASure
[:PRESsure] [z]?
Returns the pressure from the active sensor, or active
sensor if [z] is omitted.
:TEMPerature[z]?
Returns the temperature in deg. C. from the specified sen-
sor, or active sensor if [z] is omitted.
:RATE[z]?
Returns the pressure rate from the specified sensor, or ac-
tive sensor if [z] is omitted.
:BAROmetric?
Returns the barometric pressure, if a barometer sensor is
installed.
CALibration
:MODE?
Returns 1 if the sensor has been calibrated or 0 if not.
:DATE? Returns the stored calibration date.
:DATE
Sets the calibration date (YYYY,MM,DD).
:ZERO? Returns current zero point correction.
:ZERO
Sets the zero so the currently measured pressure matches
the commanded value. Limited to +/-5% FS.
ZERO:RUN
Vents the instrument and sets zero so the currently mea-
sured pressure = 0.
SENSE
[:PRESsure] [z] Selects the sensor to read or modify or active sensor if [z]
is omitted.
:NAME? Returns the accuracy and range of the sensor.
:MODE?
Returns pressure type “ABSOLUTE” or “GAUGE”
:MODE ABS/GAUGE
Sets the pressure type to absolute or gauge.
:ABS?
Returns native sensor type “1” for absolute, ”0” for gauge.
:RESolution?
Returns the FS-related relative resolution of the sensor.
:RANGE
[:UPPer]?
Returns the maximum range of the primary sensor.
:LOWer?
Returns the minimum range of the primary sensor.
:Unit
[:NAME]?
Returns a string representing the native units of the sen-
sor.
:VALue? Returns the conversion factor of the native units of the
sensor.
:REFerence
[:HEIGht]
Sets the head height correction [mm].
:MODE?
Returns head correction mode - “OFF”, “GAS”, LIQUIT”.