WIKA CPC6000 User Manual
Page 66

Automated Pressure Calibrator
CPC 6000
66
Mensor/WIKA Operating Instructions - CPC 6000
-313
Calibration data not found
-315
Configuration data not found
-350
Errorqueue overflow
-410
Query interrupted
600
Default configuration not found
601
Calibration mode active! Deactivate before setting C0..C3
602
Sensor not available
701
DCS instance not available
702
Create DCS instance failed
703
DCS still active
704
Command currently not allowed
7.4.7.4 DPI 510 Emulation Commands
The CPC 6000 will accept and respond to Druck Model DPI 510 remote commands over a GPIB as
listed below. These commands can be sent individually, or ganged together. Delimiters (separators)
between commands are not required.
Commands supported by the CPC 6000:
case ‘@’:
/*Error Status (on/off)*/
usage “@0”
sets error checking off
usage “@1”
sets error checking on
case ‘C’:
/*Control_Mode / Measure_Mode*/
usage “C0”
go to measure mode
usage “C1”
go to control mode
usage “C2”
go to standby mode
case ‘D’:
/*Display_Mode*/
usage “D0”
returns active reading in any mode
usage “D1”
returns setpoint only if in control mode, otherwise it returns 0.000
case ‘E’:
/*Clear_Error_Mode*/
usage “E “
The PCS 400 will accept all types of termination; CR/LF/EOI. It will
also accept the strings E0, E1 and E2, but will not respond to these.
Note: Send “E?” to clear an error from the buffer (this is an added function, not part of
the Druck command set).
case ‘F’:
/*Function*/
usage “F00”
does not apply to PCS 400
usage “F01”
puts instrument into vent mode when C0 or controller is off
usage “F02”
does not apply to PCS 400
usage “F03”
does not apply to PCS 400
case ‘I’:
/*IO Service Request*/
usage “I0”
dummy function accepts the string and returns what was entered