Mensor/WIKA Operating Instructions - CPC 6000
16
BR?
Returns the B channel rate.
17
BRS?
Returns the B channel rate stable flag.
18
BS?
Returns the B channel stable flag.
19
Caldisable
{YES or NO}
Disables the zero and span commands (default
= YES). When the cal-disable is YES, the zero
and span commands are disabled.
20
Caldisable?
Returns whether or not calibration of the active
sensor is disabled.
21
Cerr
none
Clears the error queue.
22
Chan
{A or B}
Sets the active channel on the instrument.
23
Chan?
Returns which channel is active.
24
Chanfunc
cr>
Sets the alternate function mode of the active
channel.
25
Chanfunc?
Returns the alternate function mode of the
channel.
26
Control
none
Instrument placed in Control mode.
27
Control?
Returns YES if active channel is in Control
mode; NO if otherwise.
28
Crate
{Slow, Medium, Fast, Vari-
able} for mid or high range
regulator, {Slow or Fast} for
low range regulator.
Sets the control slew rate for the current chan-
nel. For Variable rate see: Rsetpt command.
29
Crate?
Returns the current rate - value corresponds to
the parameters for the CRATE command.
30
Decpt
Sets the number of significant digits displayed.
31
Decpt?
Returns the number of significant digits dis-
played for the active channel.
32
Default
none
Sets the default values. This command takes
approximately 20 seconds.
33
DIO
{0 or 2}
0 turns the digital IO output pin off or low.
2 turns the digital IO output pin on or high.
34
DIO?
{value}
0 if input is low and output is low.
1 if input is high and output is low.
2 if input is low and output is high.
3 if input is high and output is high.
35
DOC
Sets the date of cal for the active sensor and
turndown.
36
DOC?
Returns the date of cal for the active sensor and
turndown.
37
DOM?
Returns the date of manufacture.
38
Error?
Returns a description of an error.