Automated Pressure Calibrator
CPC 6000

52

Mensor/WIKA Operating Instructions - CPC 6000

16

BR?

{value}

Returns the B channel rate.

17

BRS?

{YES or NO}

Returns the B channel rate stable flag.

18

BS?

{YES or NO}

Returns the B channel stable flag.

19

Caldisable

{YES or NO}

Disables the zero and span commands (default
= YES). When the cal-disable is YES, the zero
and span commands are disabled.

20

Caldisable?

{YES or NO}

Returns whether or not calibration of the active
sensor is disabled.

21

Cerr

none

Clears the error queue.

22

Chan

{A or B}

Sets the active channel on the instrument.

23

Chan?

{A or B}

Returns which channel is active.

24

Chanfunc

{Press,peak,rate,baro}<
cr>

Sets the alternate function mode of the active
channel.

25

Chanfunc?

{string}

Returns the alternate function mode of the
channel.

26

Control

none

Instrument placed in Control mode.

27

Control?

{YES or NO}

Returns YES if active channel is in Control
mode; NO if otherwise.

28

Crate

{Slow, Medium, Fast, Vari-
able} for mid or high range
regulator, {Slow or Fast} for
low range regulator.

Sets the control slew rate for the current chan-
nel. For Variable rate see: Rsetpt command.

29

Crate?

{string}

Returns the current rate - value corresponds to
the parameters for the CRATE command.

30

Decpt

{4, 5 or 6}

Sets the number of significant digits displayed.

31

Decpt?

{value}

Returns the number of significant digits dis-
played for the active channel.

32

Default

none

Sets the default values. This command takes
approximately 20 seconds.

33

DIO

{0 or 2}

0 turns the digital IO output pin off or low.
2 turns the digital IO output pin on or high.

34

DIO?

{value}

0 if input is low and output is low.
1 if input is high and output is low.
2 if input is low and output is high.
3 if input is high and output is high.

35

DOC

Sets the date of cal for the active sensor and
turndown.

36

DOC?

{mm/dd/yy}

Returns the date of cal for the active sensor and
turndown.

37

DOM?

{mm/dd/yyyy}

Returns the date of manufacture.

38

Error?

{string}

Returns a description of an error.