Duration (sec) – FUTEK IHH500 Series User Manual

Page 79

background image


EM1001-B

- 79 -


10 Thomas, Irvine, CA 92618 USA

Tel: (949) 465-0900

Fax: (949) 465-0905

Toll Free: (800) 23 FUTEK

6.3.2

Duration (SEC)

Scroll to “

DURATION

” from “

DATA LOGGING

” using the

▲▼

keys and press ENTER.

Use the

◄►

and the

▲▼

keys to select the desired duration in seconds; therefore, from 0 to 9999 seconds

can be selected.

Note: The duration that can be set is limited by the following formula:

Max Duration (sec) = Integer (21845/ sampling rate)


SPS

5

10

25

30

50

60

100

200

Max D 4369

2184

873

728

436

364

218

109


SPS

400

600

800

960

1200

1600

2400

4800

Max D 54

36

27

22

18

13

9

4

For example:
If the sampling rate has been set to 10 Samples per Second (SPS) the duration, in second, can be selected
up to 2184.
In the case of HALF SPEED the duration time can be twice as long.

If the device has been set to 10 Samples per Second (SPC) and HALF SPEED logging rate is selected, the
duration, in second, can be selected up to 4368.







SELECT DESIRED

VALUE

USING ARROW KEYS

[0000]





LOGGING RATE

► DURATION(SEC)

ACTION

------------