beautypg.com

Main display, Welcome message will be displayed at power on – FUTEK IHH500 Series User Manual

Page 15

background image


EM1001-B

- 15 -


10 Thomas, Irvine, CA 92618 USA

Tel: (949) 465-0900

Fax: (949) 465-0905

Toll Free: (800) 23 FUTEK

5.3 Main Display

Welcome message will
be displayed at power on

Auto calibration will be applied when device is
turned on. The calibration will minimize any
effects of the temperature drift for either zero
(offset) or span. During the auto calibration process
the key pad is locked out and inaccessible. Brief
information about the active channel will be
displayed when calibration is complete.


Depending on whether TEDS detect is enabled or not (see section 6.8.4), the following Sensor Profile will
be displayed for 10 seconds: TEDS enabled (left), TEDS disabled (right).






CHANNEL

01

SERIAL

1

SENSOR

BRIDGE

+2.00000

mV/V

TEMPLATE

33

SERIAL

123456

20.0000

LBS

1.98765

mV/V

AUTO CALIBRATION

IN PROGRESS

PLEASE WAIT 11

....

FUTEK ADVANCED

SENSOR TECH INC.

IPM650-PRO