Sensitivity (-), Auto calibration in progress please wait 15, Loading profile in progress please wait 30 – FUTEK IHH500 Series User Manual
Page 41
EM1001-B
- 41 -
10 Thomas, Irvine, CA 92618 USA
Tel: (949) 465-0900
Fax: (949) 465-0905
Toll Free: (800) 23 FUTEK
Note: For voltage sensor device asks for voltage output.
6.1.3.6 Sensitivity (-)
Scroll to “
SENSITIVITY (-)
” from “
NEW CHANNEL
” using the
▲▼
keys and press ENTER to set the
sensitivity for a new channel. “
SENSITIVITY
” can also be selected from “
EDIT CHANNEL
” if to make
modifications on the existing channels except channel one. Sensitivity of the channel can be selected in
mV/V using the
◄►
and the
▲▼
keys.
If the actual sensitivity is not known, but the specific range is known, use the larger number. For example
if the sensitivity of the sensor is between 2 mV/V and 3 mV/V, 3mV/V should be entered. A live
calibration would then be needed to find the actual mV/V. Any time that a new sensor sensitivity is set,
the device may do an auto calibration after it loads the new profile.
Note that sensitivity can be defined for bridge sensors and voltage sensors. This option is disabled for
current and pulse sensors.
Note: If a Uni-directional sensor is configured in “DIRECTION” the section is not available, otherwise
if the sensor is Bi-directional configured and the following step shall be taken for the negative values:
AUTO CALIBRATION
IN PROGRESS
PLEASE WAIT 15
………
LOADING PROFILE
IN PROGRESS
PLEASE WAIT 30
SELECT SENSOR
mV/V OUTPUT
USING ARROW KEYS
[2.00000]
SENSOR CONFIG
► SENSITIVITY(-)
UNIT SELECTION
SENSOR CAPACITY