Appendix a (list of probable errors), 7 appendix a (list of probable errors) – FUTEK IHH500 Series User Manual
Page 119
EM1001-B
- 119 -
10 Thomas, Irvine, CA 92618 USA
Tel: (949) 465-0900
Fax: (949) 465-0905
Toll Free: (800) 23 FUTEK
7 Appendix A (List of Probable Errors)
Error No.
Error Description on the LCD
Error #1
TEDS DEVICE WAS NOT DETECTED
Error #2
UNKNOWN DEVICE WAS DETECTED
Error #3
UNKNOWN TEMPLATE WAS DETECTED
Error #4
UNKNOWN UNIT WAS DETECTED
Error #5
ELECTRICAL VALUE IS INVALID
Error #6
TEDS DATA IS INVALID
Error #7
BRIDGE TYPE IS INVALID
Error #1 : TEDS DEVICE WAS NOT DETECTED
If any non TEDS sensor is connected or if TEDS is not connected properly the mentioned error will occur.
Error #2 : UNKNOWN DEVICE WAS DETECTED
The IHH500 supports chip number DS2430, DS2431, DS2432 and DS2433. Any other TEDS device with a
different data sheet numbers will not be able to interact with the IHH500.
Error #3: UNKNOWN TEMPLATE WAS DETECTED
The IHH500 supports template numbers 30 and 33; any other TEDS device with different template
numbers will not be able to interact information with the IHH500.
Error #4: UNKNOWN UNIT WAS DETECTED
If the sensor’s TEDS has been defined with units other than IEEE standard units, the mentioned error will
occur.
Error #5 : ELECTRICAL VALUE IS INVALID
This error occurs when the electrical values are more than defined values.
Error #6 : TEDS DATA IS INVALID
The defined TEDS information is invalid.
Error #7: BRIDGE TYPE IS INVALID
The IHH 500 supports full bridge only configurations, therefore any other bridge type transducers like
quarter or half plugged to this device will not recognized and the mentioned error to occur.