7 Appendix A (List of Probable Errors)

Error No.

Error Description on the LCD

Error #1

TEDS DEVICE WAS NOT DETECTED

Error #2

UNKNOWN DEVICE WAS DETECTED

Error #3

UNKNOWN TEMPLATE WAS DETECTED

Error #4

UNKNOWN UNIT WAS DETECTED

Error #5

ELECTRICAL VALUE IS INVALID

Error #6

TEDS DATA IS INVALID

Error #7

BRIDGE TYPE IS INVALID


Error #1 : TEDS DEVICE WAS NOT DETECTED
If any non TEDS sensor is connected or if TEDS is not connected properly the mentioned error will occur.

Error #2 : UNKNOWN DEVICE WAS DETECTED
The IHH500 supports chip number DS2430, DS2431, DS2432 and DS2433. Any other TEDS device with a
different data sheet numbers will not be able to interact with the IHH500.

Error #3: UNKNOWN TEMPLATE WAS DETECTED
The IHH500 supports template numbers 30 and 33; any other TEDS device with different template
numbers will not be able to interact information with the IHH500.

Error #4: UNKNOWN UNIT WAS DETECTED
If the sensor’s TEDS has been defined with units other than IEEE standard units, the mentioned error will
occur.

Error #5 : ELECTRICAL VALUE IS INVALID
This error occurs when the electrical values are more than defined values.

Error #6 : TEDS DATA IS INVALID
The defined TEDS information is invalid.

Error #7: BRIDGE TYPE IS INVALID
The IHH 500 supports full bridge only configurations, therefore any other bridge type transducers like
quarter or half plugged to this device will not recognized and the mentioned error to occur.