2 keypad – Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 7

Instruction Manual PC 300
3
2.2 Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed,
has a corresponding graphic indicator on the LCD. See Figure 1. Some buttons have several
functions depending on its mode of operation.
Key Function
ON/OFF
Powers on and shuts off the meter. When you switch on the meter, the meter starts up
in the mode that you last switched off from. For example, if you shut the meter off in
conductivity measurement mode, the meter will be in conductivity measurement mode
when you switch the meter on.
HOLD
Freezes the measured reading. To activate, press HOLD while in measurement mode.
To release, press HOLD again.
MODE
Selects the measurement parameter. Press MODE to toggle between pH or TDS or
conductivity measurement mode.
CAL/MEAS
Toggles between Calibration and Measurement mode.
1.
If you are in pH measurement mode, press CAL/MEAS to enter pH calibration
mode.
2.
If you are in conductivity measurement mode, press CAL/MEAS to enter
conductivity calibration mode.
3.
If you are in TDS measurement mode, press CAL/MEAS to enter TDS
measurement calibration mode.
While in SETUP sub-menu, pressing CAL/MEAS once takes you out to the SETUP
main menu and pressing CAL/MEAS second time takes you directly into the
measurement mode.
ENTER /
RANGE
ENTER function: Press to confirm values in Calibration mode and to confirm selections
in SETUP mode.
RANGE function (for conductivity &TDS measurements only): Press to enter manual
ranging function. The MEAS indicator blinks while in manual ranging function.
/ In Calibration mode:
During conductivity and TDS calibration, press to scroll through calibration values.
In SETUP mode:
Press to scroll through the setup subgroup programs.
SETUP
Takes you into the SETUP mode. This mode lets you customize meter preference and
defaults, view calibration, electrode offset data, adjust for temperature coefficient and
normalization temperature.