beautypg.com

Thermo Fisher Scientific CyberScan PC 300 User Manual

Page 15

background image

Instruction Manual PC 300

11

4.4 Conductivity

Calibration

The PC 300 meter has five measuring ranges. You can calibrate one point in each of the

measuring ranges (up to five points). If you are measuring values in more than one range,

make sure to calibrate each of the ranges you are measuring. All new calibration data will

over-ride existing stored calibration data for each measuring range you calibrate.

If you are measuring in ranges near to or greater than 20 mS (10 ppt), or

near to or lower than 100

μS (50 ppm), calibrate the meter at least once

a week to get specified ±1% F.S. accuracy.

If you are measuring in the mid ranges and you washed the probe in

deionized water and stored it dry, calibrate the meter at least once a

month.

If you take measurements at extreme temperatures, calibrate the meter

at least once a week.

Preparing for conductivity calibration

For best results, select a standard value close to the sample value you are measuring.

Alternatively use a calibration solution value that is approximately 2/3 the full-scale value of

the measurement range you plan to use. For example, in the 0 to 1999

μS conductivity range,

use a 1413

μS solution for calibration.

Table 1 - Range Indicator and its corresponding ranges

Range

indicator

Conductivity

Range

Recommended Calibration

Solution Range

TDS Range

Recommended Calibration

Solution Range

r 1

0.00 Æ 19.99

μS

6.00 to 17.00

μS

0.00 Æ 9.99 ppm

3.00 to 8.50 ppm

r 2

0.0 Æ 199.9

μS

60.0 to 170.0

μS

10.0 Æ 99.9 ppm

30.0 to 85.0 ppm

r 3

0 Æ 1999

μS

600 to 1700

μS

100 Æ 999 ppm

300 to 850 ppm

r 4

0.00 Æ 19.99 mS

6.00 to 17.00 mS

1.00 Æ 9.99 ppt

3.00 to 8.50 ppt

r 5

0.0 Æ 199.9 mS

60.0 to 170.0 mS

10.0 Æ 200 ppt

30.0 to 170 ppt

Temperature Coefficient: These meters are factory set to a temperature coefficient of 2.1%

per

°C. For most applications this will provide good results. See Program P8.1 in

Section 7.9 to set the temperature coefficient to different value. See Addendum 3,

“Calculating Temperature Coefficients” to determine the appropriate temperature

coefficient for your solution.

Normalization Temperature: The factory default value for normalization temperature is 25

°C. If you need to normalize to a value other than 25 °C, see Program P8.2 in Section
7.9.