Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 56

Instruction Manual PC 300
52
8.2 Conductivity
Electrode
Keep the conductivity probe clean. Rinse the probe twice,
and gently swirl it while you take readings. For best
accuracy, soak a dry probe for at least 5 to 10 minutes or
longer before calibration. Rinse the probe with de-ionized
water before storing. Never scratch the bands with a hard
substance. Do not strike the probe against any hard
surface.
Do not immerse the probe in oily solutions. Clean the
electrode thoroughly by stirring it in a mild detergent bath
or isopropyl alcohol. Wipe the probe with a soft tissue
paper. Rinse thoroughly in tap water and then in de-
ionized water. Recalibrate the meter after cleaning the
probe.
The conductivity probe (Order Part No. EC-
CONSEN91W/ 35608-50) which is included with your
meter features a removable probe guard to make
cleaning easy.
To remove probe guard:
1.
Grip yellow probe guard and twist clockwise. The
locking notch will release.
2.
Slide probe guard off end of probe.
Figure 43 - Conductivity probe,
EC-CONSEN91W/ 35608-50