Instruction Manual PC 300

49

7.9.1

P9.0: Resetting to factory default settings (conductivity)

Program 9.0 lets you reset all parameters to factory

default settings. This clears all calibration data and

any other conductivity setup functions you might

have changed.

From measurement mode

1. Press

the

MODE key to select conductivity

measurement mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through

subgroups until you view parameter P9.0.

4. Press

the

ENTER key to enter parameter P9.0.

5. Press

the or

keys to toggle between NO

and YES.

NO retains current settings; YES resets

to factory default settings.

6. Press

the

ENTER key to confirm selection and

to return to the measurement mode. Otherwise

press

CAL/MEAS key to return to

measurement mode without resetting to factory

default.

See Figure 42.

Notes:

To clear all pH data, see Section 7.5.

See Addendum 4 for a table of factory default settings.

Figure 42 – P9.0: Reset to factory

default values