Instruction Manual PC 300

13

NOTES:

When entering calibration mode, the meter will

display the factory default value. If the meter

was previously calibrated, the display may

“jump” to the factory default value when

switching from measurement to calibration

mode.

To exit from Conductivity calibration mode

without confirming calibration, DO NOT

press the ENTER key in step 7. Press

CAL/MEAS instead. This will retain the

meter’s old calibration data in the measuring

range of the calibration.

You can offset the conductivity reading up to

± 40% from default setting. If your measured

value differs by more than ± 40%, clean or

replace probe as needed.

Eutech Instruments/ Oakton Instruments offers a wide selection of high-quality calibration

standards. See section on Accessories for more information.

Figure 9 - Proper Immersion of

the conductivity probe