beautypg.com

P 3.3 – Thermo Fisher Scientific CyberScan PC 300 User Manual

Page 67

background image

Instruction Manual PC 300

63

17 ADDENDUM 5: SELECTING USA OR NIST BUFFER

S E T I N P 3. 3

As mentioned in Section 7.4: P3.3, program P3.3 lets

select between USA and NIST buffer sets:

USA: pH 1.68, 4.01, 7.00, 10.01, 12.45

NIST: pH 1.68, 4.01, 6.86, 9.18, 12.45

Factory default is USA buffer set.

From measurement mode

1. Press

the

MODE key to select pH measurement

mode.

2. Press

SET key to enter Setup mode.

3. Press

the or

key to scroll through subgroups

until you view parameter P3.0.

4. Press

ENTER three times to select parameter P3.3.

5. Press

the or

keys to toggle between USA and

NIST buffer sets.

6. Press

ENTER key to confirm selection and to return

to the subgroup menu. Press CAL/MEAS key to

return to measurement mode.

See Figure 44.

SETUP

usa

p 3.3

SETUP

p 3.3

Figure 44: Selecting USA or NIST

buffer set