Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the

calibration, and eventually the accuracy of the measurements. Use fresh calibration solution

each time you calibrate your meter.

All new calibration data will over-ride existing stored calibration data for each measuring range

calibrated.

Calibrating for Conductivity:

1. If necessary, press the MODE key to select

conductivity mode.

2.

Rinse the probe thoroughly with de-ionized water

or a rinse solution, then rinse with a small amount

of calibration standard.

3. Dip the probe into the calibration standard.

Immerse the probe tip beyond the upper steel

band. Stir the probe gently to create a

homogeneous sample. See Figure 8.

4. Wait for the measured conductivity value to

stabilize. If the READY indicator has been

activated (SETUP program P7.1), the READY

annunciator lights when the reading is stable.

5. Press

CAL/MEAS to enter conductivity or TDS

calibration mode. The CAL indicator will appear in

the upper right corner of the display.

6. Press

the or

keys to change the value on the

primary display to match the value of the

calibration standard.

7. Press

ENTER to confirm calibration value. The

meter returns to the MEAS (measurement) mode.

8.

Repeat steps 1 to 7 for other measuring ranges.

See figure 8.

Figure 8 - Conductivity Calibration