Instruction Manual PC 300

14

4.5 TDS

Calibration

4.5.1

Calibrating for TDS directly

The factory default setting for TDS conversion factor is

0.5. If your solution has a different TDS factor, you can

improve calibration accuracy by setting the TDS factor

prior to calibration. See P7.4 for directions.

1.

If necessary, press the MODE key to select TDS

mode.

2.

Rinse the probe thoroughly with de-ionized water or

a rinse solution, then rinse with a small amount of

calibration standard.

3.

Dip the probe into the calibration standard. Immerse

the probe tip beyond the upper steel band. Stir the

probe gently to create a homogeneous sample.

Allow time for the reading to stabilize.

4. Press

CAL/MEAS to enter TDS calibration mode.

The CAL indicator will appear in the upper right

corner of the display.

5. Press

the or

keys to change the value on the

primary display to match the value of the calibration

standard.

6. Press

ENTER to confirm the calibration value. The

meter returns to the MEAS (measurement) mode.

See Figure 10.

7.

Repeat steps 1 to 6 for other measuring ranges.

NOTES

To exit from TDS Calibration mode without confirming

calibration, DO NOT press the ENTER key in step 6.

Press CAL/MEAS instead. This will retain the meter’s old

calibration data in the measuring range of the calibration.

You can offset the TDS reading up to ± 20% from the

default setting. If your measured value differs by more

than ± 40%, clean or replace probe as needed.

Figure 10 - TDS Calibration