Instruction Manual PC 300

17

6.

Scroll with the

and

keys to set the correct temperature value (i.e. the temperature of

the temperature bath). You can adjust the reading increments of 0.1 °C.

7. Once you have selected the correct temperature press the ENTER key. The meter

automatically returns to pH measurement mode. See Figure 11.

NOTES:

You can offset the temperature reading up to ±5 °C from original reading.

To exit this program without confirming the temperature calibration value,

DO NOT press ENTER. Press CAL/MEAS instead.