Instruction Manual PC 300

61

B. Press

the key until the lower display reads P8.0.

C. Press

the

ENTER key twice. The lower display reads tCO and the upper display

shows the temperature coefficient value (should be 0, as per step 1 above).

D. Press

the key until the upper display shows your calculated temperature

coefficient.

E. Press

ENTER key to confirm the value.

F. Press

CAL/MEAS key twice to return to measurement mode.