Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 59

Instruction Manual PC 300
55
1 1 S P E C I F I CAT I O N S
Mode pH
Temperature
Conductivity
TDS
Range
-2.00 to 16.00 pH
-10.0 to 110.0 °C
( 14.0 to 230.0
°F)
0 to 19.99 µS
0 to 199.9 µS
0 to 1999 µS
0 to 19.99 mS
0 to 199.9 mS
0 to 9.99 ppm
0 to 99.9 ppm
0 to 999 ppm
0 to 9.99 ppt
0 to 99.9 ppt
Max of 199.9 ppt
based on factor
setting
Resolution
0.01 pH
0.1 °C
(0.1 °F)
0.01 µS
0.1 µS
1 µS
0.01 mS
0.1 mS
0.01 ppm
0.1 ppm
1 ppm
0.01 ppt
0.1 ppt
Accuracy
± 0.01 pH
± 0.5 °C
(±0.5 °F)
±1% Full Scale +
1 digit
±1% Full Scale +
1 digit
Calibration
Up to five points with
automatic buffer
recognition
Offset in 0.1 °C
increments
Up to five points
(one point per
range)
Up to five points
(one point per
range)
pH Slope & Offset
Display
Yes
pH Buffer Option
USA (pH 1.68, 4.01,
7.00, 10.01, 12.45)
NIST (pH 1.68, 4.01,
6.86, 9.18, 12.45)
Conductivity Cell
constant (k)
1.0
1.0
Conductivity
Temperature
Coefficient
0.0 to 10.0 % per
°C
0.0 to 10.0 % per
°C
Normalization
Temperature
15.0 to 30.0 °C
(adjustable)
15.0 to 30.0 °C
(adjustable)
Auto-ranging
Yes
Yes
Inputs
BNC
6-pin round
connector
6-pin round
connector
6-pin round
connector
Temperature
Compensation
Automatic / Manual from 0 to 100 °C
Operating
Temperature
0 to 50 °C
HOLD function
Yes
Averaging/Stability
(READY)/Auto-hold
Yes
Display Custom
Dual
LCD
Auto Power off
20 minutes after last key operation
Power
four 1.5 V AAA-sized batteries (included)
Battery life
>100 hours
Dimensions / Weight
Meter: 19 cm (L) x 10 cm (W) x 6 cm (H); 320 g
Case: 34 cm (L) x 40 cm (W) x 10 cm (H); 2.2 kg