Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 14

Instruction Manual PC 300
10
6. Press
ENTER to confirm calibration. The meter is
now calibrated to the current buffer. The lower
display scrolls through the remaining buffer options.
•
If you are performing multi-point calibration, go
to step 7.
•
If you are performing one-point calibration, go
to step 9.
7.
Rinse the electrode with de-ionized water or in rinse
solution, and place it in the next pH buffer.
8.
Follow steps 5 to 8 for additional calibration points.
See Figure 6.
9.
When calibration is complete, press CAL/MEAS to
return to pH measurement mode.
NOTES: To exit from pH calibration mode without
confirming calibration, DO NOT press ENTER in
step 6. Press CAL/MEAS instead.
If the selected buffer value is not within
±1.0 pH from
the measured pH value: the electrode and buffer icon
blink and the ERR annunciator appears in the lower left
corner of the display. See Figure 7.
To limit the number of pH buffer values available during
calibration, see SETUP section P3.2.
Figure 6 - Next point calibration for
pH 4.01
Figure 7 - Err message and
electrode icon will appear if
incorrect buffer are used