Thermo Fisher Scientific CyberScan PC 300 User Manual

Page 14

background image

Instruction Manual PC 300

10

6. Press

ENTER to confirm calibration. The meter is

now calibrated to the current buffer. The lower

display scrolls through the remaining buffer options.

If you are performing multi-point calibration, go

to step 7.

If you are performing one-point calibration, go

to step 9.

7.

Rinse the electrode with de-ionized water or in rinse

solution, and place it in the next pH buffer.

8.

Follow steps 5 to 8 for additional calibration points.

See Figure 6.

9.

When calibration is complete, press CAL/MEAS to

return to pH measurement mode.

NOTES: To exit from pH calibration mode without

confirming calibration, DO NOT press ENTER in

step 6. Press CAL/MEAS instead.

If the selected buffer value is not within

±1.0 pH from

the measured pH value: the electrode and buffer icon

blink and the ERR annunciator appears in the lower left

corner of the display. See Figure 7.

To limit the number of pH buffer values available during

calibration, see SETUP section P3.2.

Figure 6 - Next point calibration for

pH 4.01

Figure 7 - Err message and

electrode icon will appear if

incorrect buffer are used