P4.0: Resetting to factory default settings (pH)

This program lets you reset all pH parameters to factory

default settings. This clears all calibration data and any

other pH setup functions you might have changed.

The following settings will remain as you have set them:

Temperature unit of measure (°C or °F)

The temperature offset calibration value

All conductivity calibration data and parameters

From measurement mode

1. Press

the

MODE key to select pH measurement

mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P4.0.

4. Press

the

ENTER key to enter parameter P4.0. See

Figure 32.

5. Press

the or

keys to toggle between NO and

YES.

NO retains current settings

YES resets to factory default settings.

6. Press

the

ENTER key to confirm selection and to return to the measurement mode.

Otherwise press CAL/MEAS key to return to measurement mode without resetting to

factory default.

Notes:

To clear all conductivity & TDS data, see Section 7.9.1.

See Addendum 4 for a table of factory default settings.

Figure 32 – P4.0: Reset to factory

default values (pH 300 meter

only)