Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 35
Instruction Manual PC 300
31
7.1
Advanced SETUP Mode Overview
1.
In either pH or Conductivity measurement mode, press the SETUP key to enter Set up
mode.
2. Press
the or
keys to scroll through sub groups.
3. Press
ENTER key to enter a particular parameter.
See Addendum 4 for a table of meter factory default settings.
SETUP in pH Measurement Mode
P1.0: View previous calibration data
P1.1
First calibration point (pH 1.68)
P1.2
Second calibration point (pH 4.01)
P1.3
Third calibration point (pH 7.00 or 6.86)
P1.4
Fourth calibration point (pH 10.01 or 9.18)
P1.5
Fifth calibration point (pH 12.45)
Instructions on Section 7.2.
P2.0: View electrode data
P2.1
pH electrode offset
P2.2
pH electrode slope
Instructions in Section 7.3.
P3.0: Unit configuration
P3.1
READY indicator and auto endpoint function
– select on or off
P3.2
Number of pH calibration points: 2, 3, 4 or 5
P3.3
Select buffer group: USA or NIST
P3.4
Select °C or °F
Instructions in Section 7.4.
P4.0: Reset to factory defaults
P4.0
Reset meter to factory defaults
Instructions in Section 7.5.