P8.2 Adjusting the normalization temperature

Your meter will normalize its conductivity

measurements to a standard temperature that you can

select. You can adjust the normalization temperature

from 15 to 30

°C (59 to 86 °F). Meter default is 25 °C

(77

°F).

From measurement mode

1. Press

the

MODE key to select conductivity

measurement mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through

subgroups until you view parameter P8.0.

4. Press

the

ENTER key three times to select

parameter 8.2. The display shows “t.nr” on the

upper display.

5. Press

the

ENTER key again. The upper display

shows the normalization temperature and the

lower display shows “t.nr”.

6. Press

the or

keys to select the

normalization temperature.

7. Press

the

ENTER key to confirm selection and to

return to the subgroup menu. Press CAL/MEAS

key to return to measurement mode.

Figure 41 - Adjusting the

normalization temperature