Instruction Manual PC 300

45

P7.3 Selecting Automatic or Manual Temperature Compensation

This feature lets you select between Automatic

Temperature Compensation (ATC) and Manual

Temperature Compensation. Meter default is ATC.

From measurement mode

1. Press the MODE key to select conductivity

measurement mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P7.0.

4. Press

the

ENTER key three times to select

parameter 7.3. The upper display shows “ATC” and

the lower display shows “P7.3”.

5. Press

the

ENTER key again. The upper display

shows “ATC” and the lower display shows “YES” or

“NO”.

6. Press

the or

keys to select the Automatic

Temperature Compensation on or off.

YES = ATC on; NO = ATC off

7. Press

the

ENTER key to confirm selection and

proceed to P7.4 next page. Press the CAL/MEAS

key to return to measurement mode.

Figure 38 - Selecting ATC on or off