Instruction Manual PC 300

21

Taking measurements with READY indicator selected on

If the READY indicator has been activated, the READY annunciator lights when the reading is

stable*. Switch the READY indicator on or off in SETUP program P3.1. See Section 7.4.

* The READY indicator appears and the reading holds until the measured value exceeds the

tolerance (

±

0.02 pH;

±

0.8 mV <400;

±

1.2 mV > 400). Then, the READY annunciator turns

off.

Taking measurements with the auto endpoint feature selected on

When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically

“HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the

HOLD key to release the reading. Switch the Auto endpoint feature on or off in SETUP

program P 3.1, see Section 7.4.