Instruction Manual PC 300

34

7.2

P1.0: Viewing previous pH calibration data

This mode lets you recall previous pH calibration data, which helps you know when to re-

calibrate your meter. This is a “view only” mode.

From measurement mode:

1. Press

the

MODE key to select pH measurement

mode if necessary.

2. Press

the

SETUP key to enter Set up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P1.0.

4. Press

the

ENTER key repeatedly to view previous

calibration data. See Figure 26.

USA NIST

P1.1 = pH 1.68

1.68

P1.2 = pH 4.01

4.01

P1.3 = pH 7.00

6.86

P1.4 = pH 10.01 9.18

P1.5 = pH 12.45 12.45

5.

When you have scrolled through all calibration data,

you will automatically return to the subgroup menu.

Press CAL/MEAS key to return to measurement

mode.

NOTES:

If there is no previous calibration data at a particular point, the primary display will show “- - -“.

Figure 26 – P1.0: View calibration

data for pH