Thermo Fisher Scientific CyberScan PC 300 User Manual

Page 41

background image

Instruction Manual PC 300

37

P3.1: READY Indicator and auto endpoint function

This program lets you select:

READY indicator on” to indicate when the reading

is stable.

READY indicator off” for faster meter response.

Auto endpoint function on. Select auto endpoint

on to “hold” the reading when it is stable for more

than 5 seconds. The display automatically freezes,

and the HOLD indicator appears on the left side of

the display. Press the HOLD key to release the

display and access other functions.

From measurement mode

1. Press

the

MODE key to select pH measurement

mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P3.0.

4. Press

the

ENTER key to select parameter 3.1.

5. Press

the or

keys to select the configuration

you require.

OFF switches the READY indicatory off.

ON switches the READY indicator on.

• ON and HOLD together switches the auto

endpoint feature on.

6. Press

the

ENTER key to confirm selection and to

proceed to step 4 of P3.2. Press the CAL/MEAS

key to return to measurement mode.

NOTES:

Meter default is set for READY indicator on and auto endpoint function off.

Figure 29 – P 3.1: Selecting

READY function)