P3.1: READY Indicator and auto endpoint function
This program lets you select:
•
“READY indicator on” to indicate when the reading
is stable.
•
“READY indicator off” for faster meter response.
•
Auto endpoint function on. Select auto endpoint
on to “hold” the reading when it is stable for more
than 5 seconds. The display automatically freezes,
and the HOLD indicator appears on the left side of
the display. Press the HOLD key to release the
display and access other functions.
From measurement mode
1. Press
the
MODE key to select pH measurement
mode.
2. Press
SETUP key to enter Set Up mode.
3. Press
the or
keys to scroll through subgroups
until you view parameter P3.0.
4. Press
the
ENTER key to select parameter 3.1.
5. Press
the or
keys to select the configuration
you require.
•
OFF switches the READY indicatory off.
•
ON switches the READY indicator on.
• ON and HOLD together switches the auto
endpoint feature on.
6. Press
the
ENTER key to confirm selection and to
proceed to step 4 of P3.2. Press the CAL/MEAS
key to return to measurement mode.
NOTES:
Meter default is set for READY indicator on and auto endpoint function off.
Figure 29 – P 3.1: Selecting
READY function)