P5.0: Viewing Previous Conductivity Calibration data

This mode lets you recall previous calibration data, which

helps you know when to re-calibrate your meter. This is a

“view only” mode.

This function applies for conductivity & TDS mode.

From conductivity or TDS measurement mode:

1. Press

the

MODE key to select conductivity or TDS

measurement mode.

2. Press

the

SETUP key to enter Set up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P5.0.

4. Press

ENTER key repeatedly to view previous

calibration data.

5. Each calibration data corresponds to each

measurement range.

P5.1 = Range 1 (0.00 – 19.99

μS or

0.00 – 9.99 ppm)

P5.2 = Range 2 (0.0 – 199.9

μS or

0.0 – 99.9 ppm)

P5.3 = Range 3 (0 – 1999

μS or

0 – 999 ppm)

P5.4 = Range 4 (0.00 – 19.99 mS or

0.00 – 9.99 ppt)

P5.5 = Range 5 (0.0 – 199.9 mS or

0.0 – 99.9 ppt)

6.

When you have scrolled through all calibration data,

you will automatically return to the subgroup menu.

Press CAL/MEAS key to return to measurement

mode.

NOTES:

If there is no previous calibration data at a particular

point, the primary display will show “- - -“.

Figure 33 - View calibration data