Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 45
Instruction Manual PC 300
41
7.6
P5.0: Viewing Previous Conductivity Calibration data
This mode lets you recall previous calibration data, which
helps you know when to re-calibrate your meter. This is a
“view only” mode.
This function applies for conductivity & TDS mode.
From conductivity or TDS measurement mode:
1. Press
the
MODE key to select conductivity or TDS
measurement mode.
2. Press
the
SETUP key to enter Set up mode.
3. Press
the or
keys to scroll through subgroups
until you view parameter P5.0.
4. Press
ENTER key repeatedly to view previous
calibration data.
5. Each calibration data corresponds to each
measurement range.
•
P5.1 = Range 1 (0.00 – 19.99
μS or
0.00 – 9.99 ppm)
•
P5.2 = Range 2 (0.0 – 199.9
μS or
0.0 – 99.9 ppm)
•
P5.3 = Range 3 (0 – 1999
μS or
0 – 999 ppm)
•
P5.4 = Range 4 (0.00 – 19.99 mS or
0.00 – 9.99 ppt)
•
P5.5 = Range 5 (0.0 – 199.9 mS or
0.0 – 99.9 ppt)
6.
When you have scrolled through all calibration data,
you will automatically return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.
NOTES:
If there is no previous calibration data at a particular
point, the primary display will show “- - -“.
Figure 33 - View calibration data
in conductivity probe