Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 26
Instruction Manual PC 300
22
5.2
Taking Conductivity or TDS Measurement
5.2.1
Automatic Temperature Compensation
For automatic temperature compensation (ATC) simply
plug the conductivity probe into the meter (see page 6 for
directions). The ATC indicator will light on the LCD.
NOTE: If the ATC indicator does not light, manual
temperature compensation may be selected in
the meter’s SETUP mode. See Program P7.3 in
Section 7.8 for directions on selecting Automatic
Temperature Compensation.
5.2.2
Manual Temperature Compensation
IMPORTANT: For manual compensation, you must deactivate the temperature probe.
Selecting Manual Temperature Compensation for conductivity
Selecting between Automatic Temperature
Compensation (ATC) and Manual Temperature
Compensation in the SETUP program P7.3 on in
Section 7.8. Meter default is ATC on.
From Conductivity or TDS measurement mode
1) Press
SETUP key to enter Set Up mode.
2) Press
the or
keys keys to scroll through subgroups until you view parameter P7.0.
See Figure 16.
3) Press
the
ENTER key three times to select parameter 7.3. The upper display shows “ATC”
and the lower display shows “P7.3”.
4) Press
the
ENTER key again. The upper display shows “ATC” and the lower display shows
“YES” or “NO”.
Figure 15 - ATC annunciator will
light up when connected to
temperature probe
Figure 16 - P7.0 Configuration Setup
for conductivity & TDS