Instruction Manual PC 300

22

5.2

Taking Conductivity or TDS Measurement

5.2.1

Automatic Temperature Compensation

For automatic temperature compensation (ATC) simply

plug the conductivity probe into the meter (see page 6 for

directions). The ATC indicator will light on the LCD.

NOTE: If the ATC indicator does not light, manual

temperature compensation may be selected in

the meter’s SETUP mode. See Program P7.3 in

Section 7.8 for directions on selecting Automatic

Temperature Compensation.

5.2.2

Manual Temperature Compensation

IMPORTANT: For manual compensation, you must deactivate the temperature probe.

Selecting Manual Temperature Compensation for conductivity

Selecting between Automatic Temperature

Compensation (ATC) and Manual Temperature

Compensation in the SETUP program P7.3 on in

Section 7.8. Meter default is ATC on.

From Conductivity or TDS measurement mode

1) Press

SETUP key to enter Set Up mode.

2) Press

the or

keys keys to scroll through subgroups until you view parameter P7.0.

See Figure 16.

3) Press

the

ENTER key three times to select parameter 7.3. The upper display shows “ATC”

and the lower display shows “P7.3”.

4) Press

the

ENTER key again. The upper display shows “ATC” and the lower display shows

“YES” or “NO”.

Figure 15 - ATC annunciator will

light up when connected to

temperature probe

Figure 16 - P7.0 Configuration Setup

for conductivity & TDS