Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 29

Instruction Manual PC 300
25
5.2.4
Taking Measurements (Conductivity or TDS)
To take readings:
1. Rinse the probe with de-ionized or distilled water
before use to remove any impurities adhering to the
probe body. Shake or air dry. To avoid contamination
or dilution of your sample, rinse probe with a small
volume of your sample liquid.
2. Press
ON to switch on meter.
3. Press
the
MODE key to select conductivity or TDS measurement mode. The MEAS
annunciator appears on the top center of the LCD. The ATC indicator appears in the
lower right hand corner to indicate Automatic Temperature Compensation.
•
NOTE: For conductivity manual temperature compensation, you must
de-activate the temperature sensor built into the conductivity probe and
set manual temperature compensation. See Section 5.2 for more
information. The ATC indicator will disappear from the display.
4.
Dip the probe into the sample.
When dipping the probe into the sample, take care to ensure that the liquid
level is above its upper steel band. Stir the probe gently in the sample to
create a homogenous sample. See Figure 9 in Section 4.4.
5.
Allow time for the reading to stabilize. Note the reading on the display. If the READY I
indicator is selected on, it will appear when the reading is stable. See P7.1 in Section 7.8
for more information.
6. Press
the
MODE key to toggle between conductivity, TDS and pH readings.
NOTES: You can use the conductivity manual ranging function to select a specific range in which
your readings will appear. See Section 5.2.5: Manual ranging for directions.
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is
stable*. Switch the READY indicator on or off in SETUP program P 7.1. See Section 7.8 for
directions.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically
“HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the
HOLD key to release the reading. Switch the Auto endpoint feature on or off in SETUP
program P 7.1, See Section 7.8 for directions.
Figure 20 - During measurement