Thermo Fisher Scientific CyberScan PC 300 User Manual

Page 23

background image

Instruction Manual PC 300

19

5.1.2

Manual Temperature Compensation (pH)

IMPORTANT: For manual compensation, you must

disconnect the temperature probe (see section 3.2: Probe

Information).

1.

Switch the meter on. Press the MODE key to select

pH mode.

2. Press

the

CAL/MEAS key to enter pH calibration

mode. The CAL indicator will appear above the

primary display.

3.

While in pH calibration mode, press the MODE key to

enter temperature calibration mode. The primary

display shows the current temperature setting and the

secondary display shows the default value 25

°C.

4. Check the temperature of your sample using an

accurate thermometer.

5. Press

the or

keys to set the temperature to the

measured value from step 4.

6. Press

ENTER to confirm the selected temperature

and to return to the pH measurement mode.

See Figure 13.

The meter will now compensate pH readings for the

manually set temperature.

NOTES:

To exit this program without confirming the manual temperature compensation value, DO NOT

press ENTER in step 6. Press CAL/MEAS instead.

Figure 13 - Manual temperature

compensation