Instruction Manual PC 300

9

To Calibrate pH:

1.

If necessary, press the MODE key to select pH

mode. The pH indicator appears in the upper right

hand corner of the display.

2.

Rinse the probe thoroughly with de-ionized water

or a rinse solution. Do not wipe the probe; this

causes a build-up of electrostatic charge on the

glass surface.

3.

Dip the probe into the calibration buffer. The end

of the probe must be completely immersed into

the sample. Stir the probe gently to create a

homogeneous sample.

NOTE: The temperature element is in the conductivity

cell. For temperature compensated readings, dip

the conductivity cell into the calibration buffer as

well.

4. Press

CAL/MEAS to enter pH calibration mode.

The CAL indicator will be shown. The primary

display will show the measured reading while the

smaller secondary display will indicate the pH

standard buffer solution.

5.

Wait for the measured pH value to stabilize. See

Figure 5.

Figure 5 - pH Calibration