Instruction Manual PC 300

43

7.8

P7.0: Conductivity or TDS Measurement Configuration

P7.1: READY indicator and auto endpoint function

This program lets you select:

READY indicator on” to indicate when the reading

is stable.

READY indicator off” for faster meter response.

Auto endpoint function on. Select auto endpoint on

to “hold” the reading when it is stable for more than 5

seconds. The display automatically freezes, and the

HOLD indicator appears on the left side of the

display. Press the HOLD key to release the display

and access other functions.

From measurement mode

1. Press

the

MODE key to select conductivity or TDS

measurement mode.

2. Press

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P7.0.

4. Press

ENTER key to select parameter 7.1.

5. Press

the or

keys to select the configuration

you require.

OFF switches the READY indicator off;

ON switches the READY indicator on.

• ON and HOLD together switches the auto

endpoint feature on.

6. Press

ENTER key to confirm selection and to

proceed to step 4 of P7.2. You can also press the

CAL/MEAS key to return back to measurement

mode.

NOTE: Meter default is set for Ready Indicator on, and auto endpoint function off.

Figure 35 - Configuring the

READY feature

Figure 36 – P 3.1: Selecting

READY function)