Instruction Manual PC 300

42

7.7

P6.0: Viewing Conductivity Probe Data

Program 6 has five “view only” options that let you check

the probe’s parameters for diagnostic purposes. These

options show you the effective cell constant for each

range. The cell constant is adjusted according to your

calibration.

This function applies for conductivity & TDS mode.

From conductivity or TDS measurement mode

1. Press

the

MODE key to select conductivity or TDS

measurement mode.

2. Press

the

SETUP key to enter Set Up mode.

3. Press

the or

keys to scroll through subgroups

until you view parameter P6.0.

4. Press

the

ENTER key repeatedly to view the

effective cell constant for each range.

P6.1 = Range 1 (0.00 – 19.99

μS or

0.00 – 9.99 ppm)

P6.2 = Range 2 (0.0 – 199.9

μS or

0.0 – 99.9 ppm)

P6.3 = Range 3 (0 – 1999

μS or

0 – 999 ppm)

P6.4 = Range 4 (0.00 – 19.99 mS or

0.00 – 9.99 ppt)

P6.5 = Range 5 (0.0 – 199.9 mS or

0.0 – 99.9 ppt)

5.

When you have scrolled through all probe data, you

will automatically return to the subgroup menu. Press

the CAL/MEAS key to return to measurement mode.

See figure 34.

Notes

Cell constants will degrade with time and usage. You

can use this feature to alert you to the need for a new

probe prior to total failure.

(The values shown above are for illustration purpose. Your meter may display different values depending on the settings &

your environmental conditions)