beautypg.com

6 displayed digits/ over-range indication – KYORITSU 6305 Instruction Manual User Manual

Page 96

background image

7.13

KEW6305


7.6 Displayed Digits/ Over-range indication

● Digits

*

Active electrical energy WP, Apparent electrical energy WS (auto-range)

: WP1/WP2/WP3, WS1/WS2/WS3 (each phase), max 6 digits
: WP, WS (total), max 6 digits

The range is automatically assigned depending on the measured value.
The decimal point and unit are changed automatically.

Unit: Wh/ VAh

When the value exceeds 999999G, segment

becomes “

“.

However the saved data is not lost.

0.0000 -

99..9999

100.000 -

999.999

1000.00 -

9999.99

10.0000 k

-

99.9999 k

100.000 k

-

999.999 k

1000.00 k

-

9999.99 k

10000.0 k

-

99999.9 k

100000 k

-

999999 k

1000.00 M

-

9999.99 M

10000.0 M

-

99999.9 M

100000 M

-

999999 M

1000.00 G

-

9999.99 G

10000.0 G

-

99999.9 G

100000 G

-

999999 G

* Elapsed time TIME

Displayed time changes with time as follows.

Elapsed time

00:00:00

~

99:59:59

hour: minute: second

100

~

999999

hour

● Over-range indication/ others

* When the input voltage and the current exceeds the max display counts,

or

mark is

displayed on the LCD. In this case, accurate measurement cannot be made.

* On the

W

range, when the P (active power) is shown by bars “- - - -”, this means that the increment

in electrical energy is insignificant.


Refer to “

6-5-2 Over-range indication/ Bar indication” in this manual.