2 limitations of saving – KYORITSU 6305 Instruction Manual User Manual
Page 92

7.9
KEW6305
NOTE
* Never set the Function switch to OFF position during a survey otherwise the survey may be lost.
* The file number becomes “001” when;
(1) when the file number has exceeded 999
(2) after system reset
* If the same file no. exists, the old one will be overwritten.
7.5.2 Limitations of saving
● Limitation of saving (before starting a survey)
In the following cases, a survey cannot be started (manually or automatically) by pressing the
START/STOP
key.
< In case data is saved to SD card >
* When 511 files have been saved to the SD card;
mark appears and further data cannot be
saved.
Some files can to be deleted via the PC, otherwise all the saved files in the SD card can be deleted
using “
Setting 19” of Section 4 in this manual.
< In case the data is saved to internal memory >
* When 4 files have been saved to the internal memory;
mark appears and further data
cannot be saved.
● Limitations of saving (during a survey)
When the capacity of SD card or internal memory has been exceeded during a survey; measurement
continues but
mark appears on the display screen and further data will not be saved.
Press the
START/STOP
key 2 sec. or more and stop the survey once.
Refer to preceding page and delete the unnecessary file.
NOTE
* For further details about the capacity of the CF card and internal memory, refer to “
Section 9: CF card/
Internal memory” in this manual.