4.5
KEW6305
“Setting 03” Clamp sensor
Selectable Current range (“Setting 04”) differs by the selected Clamp sensors.
Clamp sensor
Current range (” Setting 04 ”)
50A (M-8128)
1 / 5 / 10 / 25 / 50A / AUTO
100A (M-8127)
2 / 10 / 20 / 50 / 100A / AUTO
200A (M-8126)
4 / 20 / 40 / 100 / 200A / AUTO
500A (M-8125)
10 / 50 / 100 / 250 / 500A / AUTO
1000A (M-8124)
20 / 100 / 200 / 500 / 1000A / AUTO
3000A
300 / 1000 / 3000A
Default value (or after system reset)
500A
1 Use the
Cursor
key on the selection screen, and select “Setting 03”.
2 Press the
ENTER
key to get the instrument into setting change mode.
3 Present setting (or default value: 500A) blinks. Select the appropriate clamp sensor with the
Cursor
key, and then press the
ENTER
key after making necessary change.
NOTE:
* Accurate results may not be obtained if the Clamp sensors in use do not match the setting
done for sensor.
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)