KEW6305

4.18

Setting 16” Demand measurement cycle

Demand measurement cycle is to be used for calculating demand values.

Setting time

NO / 10 / 15 / 30 min

Default value (or after system reset)

30 min

* Demand measurement will not be performed when “NO” has been selected.

1 Use the

Cursor

key on the selection screen, and select “Setting 16”.

2 Press the

ENTER

key to get the instrument into setting change mode.

3 Present setting (or default value: 30 min.) blinks. Press the

Cursor

key and set any desired time.

4 Press the

ENTER

key after making necessary change.