Group S

Faults / Symptoms

Cause

Remedial action

S 1

No reaction from connected

Defective signalling

Check signalling instrument(s)

(A1, A2, D1, D2)

signalling instrument(s)

instrument(s) or external

or external voltage source and

voltage source

replace if necessary.

Internal power supply (E+/E-)

Check connections and cables,

is voltage source: shorted, one

change if necessary

or several pulse outputs

(see Sect. 2.6).

defective

Voltage between E+ and E-

approx. 24 V.

Check fuse F8 on the I/O PCB

and replace if necessary

(see Sect. 8.7). If instrument

still does not operate, check

fuses F.. on the I/O PCB for the

status outputs and replace if

necessary:

F2 for terminals A1 and A

F3 for terminals A2 and A

F4 for terminals D1 and D

F5 for terminals D2 and D

⊥.

If it still does not operate, one

or several pulse outputs are

defective.

Replace I/O PCB, see Sect. 8.4.

Control inputs C1 and C2

Change setting, see

are set to ”Hold outputs” or

Sect. 4.4 and 5.10,

to ”0”

Fct. 1.11 and 1.12

In addition to this, the red

Replace signal converter,

LED flashes = Fatal Error

see Sect. 8.3.

S 2

Signalling instrument(s)

”All Error” or ”Fatal Error”

Check settings in Fct. 1.07-1.10

(A1, A2, D1, D2)

is(are) constantly tripped

settings

and change if necessary,

see Sect. 4.4 and 5.9.

S 3

No reaction of connected

Terminal ”A1” not defined as

Adjust in Fct. 3.07.

(only for A1)

signalling instrument

status output

Wrong connection/polarity

Observe polarity for driver

capacity 0.1 < I

0.2 A

see

Sect.

6.3.

A1 = ”+” and A

⊥ = ”–”

S 4

Cyclic tripping of

Terminal ”A1” not defined as

Adjust in Fct. 3.07.

(only for A1)

signalling instrument

status output

Group C

Faults / Symptoms

Cause

Remedial action

C 1

No function of control inputs Wrong connection

Connect properly,

see Sect. 2.5 and 2.6.

Defective control input C or

Check connections and cables

voltage source (internal or

and change or replace if

external)

necessary. Check voltage

source. Check fuses F6 and F7

on I/O PCB and replace if

necessary.

Wrong setting of control inputs

Change, see Sect. 4.4 and 5.10.