Part C Special applications, functional checks, service, and order numbers Sect. 7.5
05/2003
IFC 110 F
75
Group S
Faults / Symptoms
Cause
Remedial action
S 1
No reaction from connected
Defective signalling
Check signalling instrument(s)
(A1, A2, D1, D2)
signalling instrument(s)
instrument(s) or external
or external voltage source and
voltage source
replace if necessary.
Internal power supply (E+/E-)
Check connections and cables,
is voltage source: shorted, one
change if necessary
or several pulse outputs
(see Sect. 2.6).
defective
Voltage between E+ and E-
approx. 24 V.
Check fuse F8 on the I/O PCB
and replace if necessary
(see Sect. 8.7). If instrument
still does not operate, check
fuses F.. on the I/O PCB for the
status outputs and replace if
necessary:
F2 for terminals A1 and A
⊥
F3 for terminals A2 and A
⊥
F4 for terminals D1 and D
⊥
F5 for terminals D2 and D
⊥.
If it still does not operate, one
or several pulse outputs are
defective.
Replace I/O PCB, see Sect. 8.4.
Control inputs C1 and C2
Change setting, see
are set to ”Hold outputs” or
Sect. 4.4 and 5.10,
to ”0”
Fct. 1.11 and 1.12
In addition to this, the red
Replace signal converter,
LED flashes = Fatal Error
see Sect. 8.3.
S 2
Signalling instrument(s)
”All Error” or ”Fatal Error”
Check settings in Fct. 1.07-1.10
(A1, A2, D1, D2)
is(are) constantly tripped
settings
and change if necessary,
see Sect. 4.4 and 5.9.
S 3
No reaction of connected
Terminal ”A1” not defined as
Adjust in Fct. 3.07.
(only for A1)
signalling instrument
status output
Wrong connection/polarity
Observe polarity for driver
capacity 0.1 < I
≤ 0.2 A
see
Sect.
6.3.
A1 = ”+” and A
⊥ = ”–”
S 4
Cyclic tripping of
Terminal ”A1” not defined as
Adjust in Fct. 3.07.
(only for A1)
signalling instrument
status output
Group C
Faults / Symptoms
Cause
Remedial action
C 1
No function of control inputs Wrong connection
Connect properly,
see Sect. 2.5 and 2.6.
Defective control input C or
Check connections and cables
voltage source (internal or
and change or replace if
external)
necessary. Check voltage
source. Check fuses F6 and F7
on I/O PCB and replace if
necessary.
Wrong setting of control inputs
Change, see Sect. 4.4 and 5.10.