beautypg.com

KROHNE IFC 110 Converter User Manual

Page 74

background image

Sect. 7.5

Part C Special applications, functional checks, service, and order numbers

74

IFC 110 F

05/2003

Group P

Faults / Symptoms

Cause

Remedial action

P 1

Totalizer connected

Wrong connection/polarity

Connect properly, see

but does not count pulses

Sect. 2.3 and 2.6, note

recommended

resistances

!

Totalizer or exernal voltage

Check connections, totalizer

source defective

and external voltage source

and replace if necessary.

Internal power supply (E+E-)

Check connections and cables,

is voltage source, shorted or

see Sect. 2.3 and 2.6.

defective pulse output

Voltage between E+and E-

approx. 24 V. If voltage is a lot

smaller, switch off the instru-

ment, eliminate the short circuit

and replace fuses F1 and F8 on

the I/O PCB if necessary. Switch

the instrument back on. If it still

does not operate, pulse output

is defective. Replace I/O PCB or

complete

electronic

unit,

see

Sect.

8.3

and/or

8.4.

Pulse output switched off or

Switch on pulse output and

wrong flow direction setting

change flow, see Sect. 5.8 and

5.13, Fct. 1.06 (P), 1.07 (A1)

and

3.02.

Fatal Error, red LED is on

Replace signal converter or

consult KROHNE Service,

having first noted down

hardware

information

and

error

status, see Sect. 7.3, Fct. 2.02.

Control input C1 or C2 is set

Change settings, see

to ”Zero outputs” and is

Sect. 5.10, Fct. 1.11 and 1.12

activated

or deactivate control input.

Terminals A1 and A^ are not

Switch on in Fct. 3.07 and set

defined as a 2nd pulse output in Fct. 1.07.

Resistance of totalizer too low Re-position jumper X4 on

for DC operation, I > 100 mA

I/O PCB to suit DC operation,

These causes only apply

to the 2nd pulse output P2,

terminal A1 !

see

Sect.

6.3.

P 2

Constant output of

Control input C1 or C2 is set

Change setting, see

totalizer pulses

to ”Hold outputs” and is

Sect. 5.10, Fct. 1.11 and 1.12

activated

or deactivate control input.

P 3

Unsteady pulse rate

Electrical conductivity of

Increase time constant

process liquid is too low

(see Sect. 6.5-6.8) or

consult

KROHNE

Service.

P 4

Pulse rate too high or too low

Incorrect pulse output settings Correct settings in

Fct. 1.06 (P) or 1.07 (A1).