Part B IFC 110 F Signal converter Sect. 5.3

05/2003

IFC 110 F

41

5.3 Low-flow

cutoff

SMU

Fct. 1.03 L.F. CUTOFF
Press

key


Select
OFF

(fixed tripping points:

ON = 0.1 % / OFF = 0.2 %)

PERCENT

(variable tripping points:

ON = 1 - 19 % / OFF = 2 - 20 %)

Select with keys

and (only if PERCENT is selected).

1st number (cursor) flashes.

Setting the numerical value when ”PERCENT” is selected
01 to 19

(cutoff ”ON” value, left of hyphen)

02 to 20

(cutoff ”OFF” value, right of hyphen)

Change flashing number (cursor) with keys

and ↓ .

Use

andkeys to shift cursor 1 place to right or left.

Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press

key to return to Fct. 1.03 L.F. CUTOFF.

Note: the cutoff ”OFF” value must be greater than the cutoff ”ON” value.

5.4 Display

Fct. 1.04 DISPLAY
Press

key

DISP. FLOW = select unit for display of flow rate, press key

NO DISP

(no display)

m³/h

(cubic metres per hour)

Liter/Sec

(litres per second)

US.Gal/min

(US gallons per minute)

• User-defined unit, factory setting = ”Liter/h” (litres per hour) or ”US MGal/day”,

see Section 5.14

PERCENT

(percentage display)

BARGRAPH

(numerical value and bar graph display in %)

Select with

andkeys.

Press

key to change to subfunction ”DISP. TOTAL.”.

DISP. TOTAL. = select unit for totalizer display, press key

NO DISP.

(no display)

OFF

(internal totalizer switched off)

+ TOTAL.

– TOTAL.

+/– TOTAL.

SUM. (Σ)

ALL (sequential)

Select with

andkeys.

Press

key to change to display unit setting.


(cubic metres)

Liter

(litres)

US.Gal

(US gallons)

• User-defined unit, factory setting = ”Liter” (litres) or ”US MGal/day”, see Section 5.14

Select with

and keys.

Use

key to change to totalizer format setting.