Part A System installation and start-up Sect. 1.3.4

05/2003

IFC 110 F

11


1.3.4 Cable lengths (max. distance between signal converter and flow sensor)

Abbreviations and explanations
The abbreviations used in the following tables, diagrams and connection diagrams stand for:
A

Signal line A (type DS) with double shielding, max. length see diagram

B

Signal line B (type BTS) with triple shielding, max. length see diagram

C

Field current line, minimum cross section (A

F

) and max. length see table

D

High-temperature silicone line, 3x1.5 mm² (14 AWG) Cu, with single shield,
max. length 5 m (16 ft)

E

High-temperature silicone line, 2 x 1.5 mm² (14 AWG) Cu, max. length 5 m (16 ft)

A

F

Cross section of field current line C in Cu, see table

L

Cable length in m or ft

Κ

Electrical conductivity of the process liquid

ZD Intermediate connection box required in connection with lines D and E for flow sensors

ALTOFLUX 4000 F, PROFIFLUX 5000 F and VARIFLUX 6000 F for process temperatures
exceeding 150°C (302°F).

Recommended length of signal line
for magnetic field frequencies

1/6 x power frequency

Flow

Meter size

Signal

Please note!

sensor

DN mm

Inch

line

VARIFLUX 6000 F

2.5 - 15

1

/

10

-

1

/

2

B1

25

- 80 1

- 3

A1 / B3

PROFIFLUX 5000 F

2.5 -

1

/

10

-

B1

4

- 15

1

/

6

-

1

/

2

B2

25

- 100 1

- 4

A1 / B3

ALTOFLUX 4000 F

10

- 150

3

/

8

- 6

A1 / B3

200

- 1200 8

- 48

A2

/

B4

ALTOFLUX 2000 F

150

- 250 6

- 10

A2 / B4

ECOFLUX 1000 F

10

- 150

3

/

8

- 6

A1 / B3

M900

10

- 300

3

/

8

- 12

A2 / B4

For application with
empty pipe detection
(EPD) max. length
< 20 m / 66 ft.

Max. length and minimum cross section of field current line
Length L

Cross section AF (Cu), minimum

0 to 150 m

5 to 500 ft 2 x 0.75 mm² Cu

2 x 18 AWG

150 to 300 m

500 to 1000 ft 2 x 1.50 mm² Cu

2 x 14 AWG

300 to 600 m

1000 to 1900 ft 4 x 1.50 mm² Cu

4 x 14 AWG