KROHNE IFC 110 Converter User Manual
Page 11
Part A System installation and start-up Sect. 1.3.4
05/2003
IFC 110 F
11
1.3.4 Cable lengths (max. distance between signal converter and flow sensor)
Abbreviations and explanations
The abbreviations used in the following tables, diagrams and connection diagrams stand for:
A
Signal line A (type DS) with double shielding, max. length see diagram
B
Signal line B (type BTS) with triple shielding, max. length see diagram
C
Field current line, minimum cross section (A
F
) and max. length see table
D
High-temperature silicone line, 3x1.5 mm² (14 AWG) Cu, with single shield,
max. length 5 m (16 ft)
E
High-temperature silicone line, 2 x 1.5 mm² (14 AWG) Cu, max. length 5 m (16 ft)
A
F
Cross section of field current line C in Cu, see table
L
Cable length in m or ft
Κ
Electrical conductivity of the process liquid
ZD Intermediate connection box required in connection with lines D and E for flow sensors
ALTOFLUX 4000 F, PROFIFLUX 5000 F and VARIFLUX 6000 F for process temperatures
exceeding 150°C (302°F).
Recommended length of signal line
for magnetic field frequencies
≤
1/6 x power frequency
Flow
Meter size
Signal
Please note!
sensor
DN mm
Inch
line
VARIFLUX 6000 F
2.5 - 15
1
/
10
-
1
/
2
B1
25
- 80 1
- 3
A1 / B3
PROFIFLUX 5000 F
2.5 -
1
/
10
-
B1
4
- 15
1
/
6
-
1
/
2
B2
25
- 100 1
- 4
A1 / B3
ALTOFLUX 4000 F
10
- 150
3
/
8
- 6
A1 / B3
200
- 1200 8
- 48
A2
/
B4
ALTOFLUX 2000 F
150
- 250 6
- 10
A2 / B4
ECOFLUX 1000 F
10
- 150
3
/
8
- 6
A1 / B3
M900
10
- 300
3
/
8
- 12
A2 / B4
For application with
empty pipe detection
(EPD) max. length
< 20 m / 66 ft.
Max. length and minimum cross section of field current line
Length L
Cross section AF (Cu), minimum
0 to 150 m
5 to 500 ft 2 x 0.75 mm² Cu
2 x 18 AWG
150 to 300 m
500 to 1000 ft 2 x 1.50 mm² Cu
2 x 14 AWG
300 to 600 m
1000 to 1900 ft 4 x 1.50 mm² Cu
4 x 14 AWG