5.20 Range
change
Automatic range change by status output
Fct. 1.07 - 1.10 Status outputs A1, A2, D1 or D2
(Define operating mode of output terminal A1, see Section 5.18)
Press
→ key
Press
↑ key as often as required to set one of the status outputs to automatic range change
”AUTO. RNG.”
Press
↵ key to change to numerical setting, 1st number (cursor) flashes.
Change flashing number (cursor) with keys
↑ and ↓. Use → and ← keys to shift cursor
1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Setting range: 5 – 80 PERCENT of Q
100%
(= low to high range ratio 1:20 to 1:1.25)
Press
↵ key to return to Fct. 1.07 - 1.10, status outputs A1, A2, D1 or D2.
External range change by control input
Fct. 1.11 or 1.12 Control inputs C1 or C2
Press
→ key
Press
↑ key as often as required to set one of the control inputs C1 or C2 to range change
”EXT. RNG.”
Press
↵ key to change to numerical setting, 1st number (cursor) flashes.
Change flashing number (cursor) with keys
↑ and ↓. Use → and ← keys to shift cursor
1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Setting range: 5 – 80 PERCENT of Q
100%
(= low to high range ratio 1:20 to 1:1.25)
Press
↵ key to return to Fct. 1.11 or 1.12, control inputs C1 or C2.