Setting of totalizer format
• Auto (exponent notation)
• # . #######
• ##### . ###
• ## . ######
• ###### . ##
• ### . #####
• ####### . #
• #### . ####
• ########
Select with
↑ and ↓ keys.
Press
↵ key to change to subfunction ”DISP. MSG.”.
→ DISP. MSG. = additional messages desired in measuring mode, press → key
• NO
(no additional messages)
• YES
(display additional messages, e.g. errors, in sequence with measured values)
Select with
↑ and ↓ keys.
Press
↵ key to return to Fct. 1.04 DISPLAY
Note: ”BUSY” is displayed in measuring mode when all displays are set to ”NO DISP.” or ”NO”.
Sequencing of displays is automatic. In measuring mode, however, keys - and ¯ can be used
for manual sequencing. Return to automatic sequencing after approx. 3 minutes.
Please refer to Section 2.7 ”Factory settings”
5.5
Internal electronic totalizer
The internal electronic totalizer counts in m³ regardless of the unit set in Fct. 1.04, subfunction
”DISP. FLOW”.
The counting range depends on the meter size and has been selected such that the totalizer will
count for at least 1 year without overflow.
Meter size
Counting range
DN mm
Inch
in m³
US Gal equivalent
2.5 - 50
1
/
10
- 2
999 999.99999999
0 - 264 172 052.35800
65 - 200
2
1
/
2
- 8
9 999 999.9999999
0 - 2 641 720 523.5800
250 - 600
10
- 24
99 999 999.999999
0 - 26 417 205 235.800
700 -1000
28
- 40
999 999 999.99999
0 - 264 172 052 358.00
Only part of the totalizer count is shown in the display as it is not possible to display a 14-digit
number. Unit and format of the display are freely selectable. Refer to Fct. 1.04, subfunction ”DISP.
TOTAL.” and Section 5.4 to determine which part of the count is to be displayed. Display overflow
and totalizer overflow are independent of one another.
Example
Internal count
0000123 . 7654321
m³
Format, display unit
XXXX . XXXX
liter
Internal count in unit
0123765 . 4321000
liter
Displayed
3765 . 4321