Part B IFC 110 F Signal converter Sect. 5.8
05/2003
IFC 110 F
45
→ PULSWIDTH = select pulse width, press → key
• AUTO
(automatic = 50 % of cycle duration of 100 % output frequency)
• SYM.
(symmetric = pulse duty factor 1:1 across entire range)
• SEC.
(variable) setting range 0.01 - 1.00 SEC
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting.
Change flashing number (cursor) with keys
↑ and ↓ . Use → and ← keys to shift cursor
1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press
↵ key to change to subfunction ”VALUE P” and/or ”VALUE P2”.
→ VALUE P = set pulse value per unit volume
only appears when ”PULSE/VOL.” is selected in ”SELECT P”, press
→ key
• XXXX PulS/m³
• XXXX PulS/litre
• XXXX PulS/US.Gal
• XXXX PulS/user-defined unit, factory setting = ”Liter” or ”US MGal/day”, see Section 5.14
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting, 1st number (cursor) flashes.
Set numerical value
• XXXX
(setting range depends on pulse width and full-scale range:
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
100%
)
Change flashing number (cursor) with keys
↑ and ↓ . Use → and ← keys to shift cursor
1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press
↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.
or
→ VALUE P = set pulse value per unit time
only appears when ”PULSE/TIME” is selected in ”SELECT P”, press
→ key
• XXXX PulSe/Sec
• XXXX PulSe/min
• XXXX PulSe/h
• XXXX PulSe/user-defined unit, factory setting = ”h” or ”day”, see Section 5.14
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting, 1st number (cursor) flashes.
Set numerical value
• XXXX
(setting range depends on pulse width)
Change flashing number (cursor) with keys
↑ and ↓ . Use → and ← keys to shift cursor
1 place to right or left.
Flashing numbers (cursor) can also be directly set with the 10-key keyboard.
Press
↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.
Please refer to Section 2.7 ”Factory settings”.
Refer to Section 2.6 for connection diagrams and to Section 5.16 for characteristics.