beautypg.com

KROHNE IFC 110 Converter User Manual

Page 34

background image

Sect. 4.4

Part B IFC 110 F Signal converter

34

IFC 110 F

05/2003

Fct. Text

Description

and

setting

1.08

STATUS A2

Status outputs A2, D1 and D2

1.09 STATUS

D1

Description of function of status outputs A2, D1 and D2

1.10 STATUS

D2

}

on the next page but one.

1.11

CONTROL C1

Countrol inputs C1 and C2

1.12 CONTROL

C2

}

Description of function of control inputs on the next page but one.

1.06

PULS P

Pulse output P for electronic totalizers up to 10,000 pulses/s

1.07

PULS2 A1

2nd pulse output A1 for electromechanical totalizers up to max. 50 Hz.

Connection of terminal A1 as a 2nd pulse output A1 or as status output A1,
see Fct. 3.07 HARDWARE, ”Terminal A1”.

FUNCT. P

Selecting the function for pulse outputs P and P2

FUNCT. P2

● OFF

● + DIR.

● - DIR. (measuring in one flow direction only)

● 2 DIR. (forward/reverse flow, F/R mode)

Press

key to change to subfunction ”SELECT P or P2”.

SELECT P

Selecting the type of pulse

SELECT P2

● PULSE/VOL. (pulses per unit volume, flow rate)

● PULSE/TIME (pulses per unit time for 100% flow rate)

Press

key to change to subfunction ”PULSWIDTH”.

PULSWIDTH

Selecting the pulse width

PULSWIDTH

● 0.01 - 1.00 s (only for F

max

< 50 pulses/s)

● AUTO (automatic = 50% of cycle duration of 100% output frequency)

● SYM (symmetric = pulse duty factor approx. 1:1 across the entire range)

Press

key to change to subfunction ”VALUE P or P2”.

VALUE P

Setting the pulse value per unit volume (only displayed when

VALUE P2

”PULSE/VOL.” is selected in ”SELECT P or P2” above).

● xxxx PulS/m³

● xxxx PulS/Liter

● xxxx PulS/US.Gal

● xxxx PulS/user unit, factory setting ”Liter” or ”US MGal” (s. Fct. 3.05)

Setting range ”xxxx” depends on pulse width and full-scale range:

P

min

= F

min

/ Q

100%

, P

max

= F

max

/ Q

100%

Press

key to return to Fct. 1.06 PULS P or Fct. 1.07 PULS2 A1.

VALUE P

Setting the pulse value per unit time (only displayed when

VALUE P2

”PULSE/TIME” is selected in ”SELECT P or P2” above).

● xxxx PulSe/Sec (=Hz)

● xxxx PulSe/min

● xxxx PulSe/h

● xxxx PulSe/user unit, factory setting ”h” (s. Fct. 3.05)

Setting range ”xxxx” depends on pulse width (see above).

Fct. 1.06 and 1.

0

7 have identical

menus and are c

onfigured in

accordance

w

ith the same setting mode.

Press

key to return to Fct. 1.06 PULS P or Fct. 1.07 PULS2 A1.