Part B IFC 110 F Signal converter Sect. 4.4
05/2003
IFC 110 F
35
Fct. Text
Description
and
setting
1.07 STATUS
A1
Status
output
A1
(terminal A1 connected as status output A1 or
as a 2nd pulse output A1, see Fct. 3.07 HARDWARE, ”terminal A1”)
1.08 STATUS
A2
Status
output
A2
1.09
STATUS D1
Status output D1
1.10
STATUS D2
Status output D2
→
● OFF
● ON
● ALL ERROR
● FATAL.ERROR
● INVERS D1 (inverse mode of D1 and D2)
● INVERS A1 (inverse mode of A1 and A2 possible only if A1
is operated as status output, see Fct. 3.07 HARDWARE, ”terminal A1”)
● SIGN I, P or P2
(F/R
mode)
● OVERFL. I, P or P2
(overloading
the
outputs)
}
dynamic behaviour
of outputs see Fct. 1.02
TIMECONST.:
I = ONLY I
P or P2 = ALL
● EMPTY PIPE (”tube empty” signal only with built-in option)
● TRIP. POINT
Press → key to change to character.
Selection:
● + DIR.
● - DIR.
● 2 DIR.
Press ↵ key to change to numerical setting.
Setting
range: 000 - 150 PERCENT
● AUTO. RNG. Setting range: 05-80 PERCENT (= lower to upper range ratio
1:20 to 1:1.25, value must be higher than that of Fct. 1.03 L.F. CUTOFF)
Fct. 1.07 to 1.1
0
are configured in
accordance
w
ith the same setting
mode. Functions
set for one of th
e
status outputs ar
e no longer available
for the oth
er stat
us outputs.
Press ↵ key to change to numerical setting.
Press ↵ key to return to Fct. 1.06, 1.07, 1.08 or 1.09.
Fct.
Text
Description and setting
1.11
CONTROL C1
Control input C1 and C2
1.12 CONTROL
C2 ● OFF
● EXT. RNG. (external range change)
→
Setting
range: 05-80 PERCENT (= lower to upper range ratio 1:20 to 1:1.25,
value must be higher than that of Fct. 1.03 L.F. CUTOFF)
Press ↵ key to change to numerical setting.
● OUTP. HOLD (hold output values)
● OUTP. ZERO (set outputs to ”min. values”)
● TOTAL.RESET (reset the totalizer)
● ERROR.RESET (delete error messages)
Press ↵ key to return to Fct. 1.11 or 1.12 CONTROL C1 or C2
Fct.
Text
Description and setting
2.0
TEST
Test menu
TEST Q
Test measuring range Q
Precautionary
query
● SURE NO
Press
↵ key to return to Fct. 2.01 “TEST Q”.
● SURE YES
Press
↵ key, then use ↑ key to
select value: -110 / -100 / -50 / -10 / 0 / +10 / +50 / +100 / +110 PCT.
of set full-scale range Q
100%
.
Displayed value is available at outputs I and P.
Press
↵ key to return to Fct. 2.01 “TEST Q”.
2.02
HARDW. INFO
Hardware information and error status
Before consulting factory, please note down all 6 codes.
→ MODUL ADC
X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵ key to transfer to “MODUL IO“.
→ MODUL IO
X . X X X X X . X X
Y Y Y Y Y Y Y Y Y Y
Press
↵ key to transfer to “MODUL DISP.“.
→ MODUL DISP.
X . X X X X X . X X
Press
↵ key to return to
Y Y Y Y Y Y Y Y Y Y
Fct. 2.02 “HARDW. INFO“.
2.03
HARDW. TEST
Hardware test (Precautionary query)
- SURE NO
Press
↵ key to return to Fct. 2.03 ”HARDW. TEST”.
-
SURE
YES Press ↵ key to start test, duration approx. 60 s
If errors are found, the first one is displayed. Press
↓ key to
display next error. List of errors see Section 4.5.
Press
↵ key to return to Fct. 2.03 ”HARDW. TEST”.